Through post-market surveillance, ELA Medical has observed that a no-output condition could occur in a limited number of Symphony or rhapsody pacemakers. The no-output condition could occur due to metal migration caused by a specific manufacturing process.
Source facts
Action
A Dear Doctor letter, dated 10/25/05, was sent to all affected doctors. The letter describes the failures and mentions that pacemaker depend patients implanted with devices manufactured in the first group could require prophylactic replacement.
Initiated
2005-10-25
Root cause
Other
Terminated
2007-01-28
Distribution
Devices distributed with the US to AZ, CA, FL, IL, LA, MA, ME, MO, MS, SC,
Recall number
Z-0267-06
Recall status
Terminated
CDRH record ID
42981
FDA classified
2005-12-02
Recall event number
34003
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.