Z-0267-2013: Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of differ

Recall record Z-0267-2013 for Medacta Usa Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0267-2013
Company in FDA record
Medacta Usa Inc
FDA classification date
Nov 8, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.

Source facts

Action
Medacta USA, sent a Urgent Safety Notification letter dated October 13, 2008, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The letter informed the customers of the lot numbers on recall and the reference numbers. The recall was initiated because the offset broach handles contain a lever mechanism that is connected to the locking axis by means of a metallic wire. The rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches. For further questions please call (858) 705-0350.
Initiated
2008-10-13
Root cause
Nonconforming Material/Component
Terminated
2012-11-09
Distribution
Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.
Recall number
Z-0267-2013
Recall status
Terminated
CDRH record ID
85292
FDA classified
2012-11-08
Recall event number
53337

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗