Open original FDA record ↗Recall record
- FDA identifier
- Z-0267-2013
- Company in FDA record
- Medacta Usa Inc
- FDA classification date
- Nov 8, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.
Source facts
- Action
- Medacta USA, sent a Urgent Safety Notification letter dated October 13, 2008, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The letter informed the customers of the lot numbers on recall and the reference numbers. The recall was initiated because the offset broach handles contain a lever mechanism that is connected to the locking axis by means of a metallic wire. The rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches. For further questions please call (858) 705-0350.
- Initiated
- 2008-10-13
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-11-09
- Distribution
- Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.
- Recall number
- Z-0267-2013
- Recall status
- Terminated
- CDRH record ID
- 85292
- FDA classified
- 2012-11-08
- Recall event number
- 53337