Z-0268-2020: VITROS chemistry Products Cl- Slides, Catalog No. 8445207 (which does NOT support Urine CL-). MicroSlides

Recall record Z-0268-2020 for Ortho-Clinical Diagnostics. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0268-2020
Company in FDA record
Ortho-Clinical Diagnostics
FDA record creation date
Nov 1, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Users may experience intermittent slide dispense issues with some cartridges due to the anti-backup platen failing to advance inside the cartridge.

Source facts

Action
On September 10, 2019, the firm distributed Urgent Product Correction Notification letters to affected customers. Customers were asked to take the following actions: "Complete and return the Confirmation of Receipt form. "If you experience a slide dispense issue, try to reposition the ABP by following the instructions in the enclosed Resolving Slide Dispense Issues document. Please place the instructions by each VITROS System in your facility that processes Microslides. "If you are unable to reposition the ABP, remove the affected cartridge and replace it with a new one. Complete and return the Request for Credit form as needed to receive credit for the slides remaining within affected and discarded cartridges. "Please forward this notification if the product was distributed outside of your facility. The firm will provide credit for the slides remaining within affected cartridge for which you are unable to reposition the ABP. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311.
Initiated
2019-09-10
Root cause
Process control
Terminated
2020-11-13
Distribution
Domestic distribution nationwide. Foreign distribution to Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, the Netherlands, Norway, Poland, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
Recall number
Z-0268-2020
Recall status
Terminated
CDRH record ID
176664
FDA record created
2019-11-01
Recall event number
83879

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗