Z-0269-06: Abbott brand CELL-DYN 1800 DILUENT Packaged in 3.8L containers, List Numbers: 99226-01, 08H17-01,(on bo

Recall record Z-0269-06 for Abbott Laboratories. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0269-06
Company in FDA record
Abbott Laboratories
FDA classification date
Dec 6, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The product failed 12-month stability testing that can lead to falsely reduced hemoglobin (Hgb) test results.

Source facts

Action
On September 20, 2005, the firm issued letters to all its direct consignees, informing them of the affected product and providing instructions on the recall. In February, the firm expanded its recall to all lots as a result of having identified an additional failure in a different lot.
Initiated
2005-09-20
Root cause
Other
Terminated
2006-11-02
Distribution
Distribution of DILUENT is worldwide to a total of 1,584 consignees. All CELL-DYN 1800 customers were notified via the Device Recall Letter. At the time of receipt of the letter, the customer may or may not have the recalled lots in their inventory. Therefore, the letter directed the customer to assess their inventory of DILUENT 08H17-01 and 99226-01 to determine if they had any of the affected lots in their possession. Foreign countries include: Canada, Columbia, Uruguay, Germany, Barbados, Bermuda, Cayman Islands.
Recall number
Z-0269-06
Recall status
Terminated
CDRH record ID
42161
FDA classified
2005-12-06
Recall event number
33616

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗