Open original FDA record ↗Recall record
- FDA identifier
- Z-0269-2008
- Company in FDA record
- Siemens Medical Solutions USA Inc
- FDA classification date
- Nov 29, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect Decay Correction Results - ECAT System's software versions 7.2.2 and 7.4 calculate the decay correction for emission data incorrectly, while performing whole-body studies with all transmissions followed by all emissions.
Source facts
- Action
- Siemans sent an Urgent Device Correction letter, dated June 15, 2007, to all affected consignees informing the user of the problem and requesting that they discontinue the practice of running all transmissions followed by all emissions until a corrective software patch is installed. The corrective software patch is expected to be installed by October 2007.
- Initiated
- 2007-07-23
- Root cause
- Software design
- Terminated
- 2009-09-11
- Distribution
- Worldwide Distribution - USA and Countries of Argentina, Austria, Belgium, Brazil, Canada, Switzerland, Chile, China, Czech Republic, Germany, Denmark, Spain, Finland, France, Great Britain, Italy, Japan, South Korea, Netherlands, Norway, Saudi Arabia, Sweden, Slovak Republic, Turkey, and Taiwan.
- Recall number
- Z-0269-2008
- Recall status
- Terminated
- CDRH record ID
- 64980
- FDA classified
- 2007-11-29
- Recall event number
- 45154