Z-0269-2014: Philips brand IntelliSpace PACS 4.4, Image Management System; The system is a software package used with

Recall record Z-0269-2014 for Philips Healthcare Informatics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0269-2014
Company in FDA record
Philips Healthcare Informatics, Inc.
FDA classification date
Nov 13, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Philips Healthcare became aware of a software problem with the IntelliSpace PACS system, in that a timing issue has been identified that may result in a condition where the device does not store additional images received before processing is complete. The missing images may contain pathology which may result in a different diagnosis if they had been available at time of interpretation.

Source facts

Action
Philips sent an Urgent Field Safety Notice dated October 14, 2013, to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Please review the following with all radiology technicians who use lntelliSpace BY CUSTOMER/USER PACS: " Assure you and your colleagues are aware of the problem. " Verify that all images sent for each study are present on the iVault before the images are purged from the modality. o Use the image count from the lower right corner of the thumbnails as seen in Enterprise only. Single images (such as CR, DX and MG) show no image count, so the user should count those thumbnails without image counts as one. Add all the numbers up to come up with the total number of viewable images, stored to lntelliSpace. " Compare number of images sent from the modality and those sent from other 3rd party source to image count in IntelliSpace PACS. " If the image count on the modality or 3 party source indicates there are more images to the study than are present on the iVault, the study should be resent from the modality and/or 3 party source. " Resending the studies will typically correct the problem but the technician should again verify the image count is correct. If you need any further information or support concerning this issue, please INFORMATION AND contact your Philips representative or contact Care at isitecare@philips.com. Please call (650) 293-2300 for further questions.
Initiated
2013-10-14
Root cause
Software in the Use Environment
Terminated
2014-06-11
Distribution
Worldwide Distribution - USA (nationwide) and country: Canada.
Recall number
Z-0269-2014
Recall status
Terminated
CDRH record ID
122607
FDA classified
2013-11-13
Recall event number
66560

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗