Skip to content

Public FDA record

Z-0269-2019: REF VV19F Reinforced Dual Lumen Cannula, 19Fr., Sterile EO, DISPOSABLE: DO NOT RE-USE CE 0473 Product U

Recall record Z-0269-2019 for OriGen Biomedical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0269-2019
Company in FDA record
OriGen Biomedical, Inc.
FDA classification date
Nov 7, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage

Source facts

Action
On April 1, 2015 the firm sent an MEDICAL DEVICE MARKET WITHDRAWAL to customers/distributors via email to inform them the catheters may start to show signs of failure at the junction of the clear extension tube and blue hub connection that can be recognized by air entrainment or blood leakage. If the preliminary failure is identified, the catheter should be replaced immediately. Serious injuries could occur due to the failure mode. The customers are instructed to discontinue use of all products from the affected lot and to return any unused products immediately and to complete and return the acknowledgement and receipt form even if the customer does not currently have products in its inventory. The firm is conducting an investigation to determine root cause of the failure and will notify the customers when the investigation has been completed and when replacement products are available. If customers have any questions regarding the return of product or the status of the investigation, they can contact: OriGen Biomedical Attn: Bernie R Silvers 7000 Burleson Road, Building D Austin, TX 78744 Phone Number +1 512-474-7278
Initiated
2015-03-30
Root cause
Mixed-up of materials/components
Terminated
2024-03-01
Distribution
Worldwide Distribution - US Nationwide in the states of: AR, AZ, CA, FL, GA, IL, IN, LA, MA, MI, MO, NC, NY, OH, OR, PA, RI, TX, WA, and WI OUS: Belgium, Canada, Germany, Kingdom of Saudi Arabia, Poland, Spain Sweden, Thailand and United Kingdom
Recall number
Z-0269-2019
Recall status
Terminated
CDRH record ID
167608
FDA classified
2018-11-07
Recall event number
81164

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗