Z-0269-2023: MicroScan autoSCAN-4 Instrument, Catalog Number B1018-280

Recall record Z-0269-2023 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0269-2023
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Nov 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to MicroScan autoSCAN-4 Instruments being manufactured with an incorrect diffuser plate.

Source facts

Action
During the week of 10/03/2022, an Urgent Medical Device Recall letter was sent via email and a hardcopy via standard postal mail. The letter is informing customers that the diffuser plates used with the MicroScan autoSCAN-4 were incorrectly manufactured with the locator notch in the upper right corner with the shiny side up. Customers are instructed to follow steps outlined in Attachment 1 of the notice: -If they confirm the diffuser plate was manufactured with the locator notch in the upper right corner with the shiny side up, follow the steps in Attachment 1, Part II to reverse this shield so that it is installed frosted side up, the notch will now be on the left. - If they confirm the diffuser plate was manufactured with the locator notch in the upper right corner with the diffuser plate frosted side up, then no further action is required. - Ensure the diffuser plate continues to be installed with frosted side up (as described in Attachment 1, Part II ) when performing maintenance each day before first use per MicroScan autoSCAN-4 Instrument Guide and until you receive replacement. - They may choose to display Attachment 1 next to your AS4 to ensure this work around is performed correctly until replacement is available. - Contact Customer Technical Support 800 677-7226 option 1 for replacement of the diffuser plate (P/N 5639-0001). - Once the replacement is received, discard the diffuser plate in use, install the new diffuser plate and create a Water Blank File with the newly placed shield to ensure optimization of the panel reads. - Check their P/N 5639-0001 inventory, if their diffuser plate was manufactured with the locator notch in the upper right corner with the shiny side up, contact Customer Technical Support 800 677- 7226 option 1 for replacement. For questions, contact Customer Technical Support at http://www.beckmancoulter.com or call 800-677-7226 option 1 or Client Services 800-526-3821
Initiated
2022-10-03
Root cause
Nonconforming Material/Component
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, FL, IA, ID, MI MS, MT, NC, NE, PA, SC, TX WA, WI, and WV. The countries of Chile, Germany, Indonesia, Italy, Japan, Mexico, Peru, Portugal, South Africa, Spain, United Kingdom of Great Britain and Northern Ireland, and Uruguay.
Recall number
Z-0269-2023
Recall status
Open, Classified
CDRH record ID
196240
FDA classified
2022-11-18
Recall event number
90941

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗