Open original FDA record ↗Recall record
- FDA identifier
- Z-0270-2013
- Company in FDA record
- Primus Medical LLC
- FDA classification date
- Nov 8, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.
Source facts
- Action
- The firm, Primus Medical LLC, sent an "Important Notification Relating to Primus Medical Product TL500" letter dated March 29, 2011 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to replace the "Retaining washers & Clevis Pins" with the "Clevis Pins & Bow Tie Cotter Pins" in the TL500-001 Service Pack provided with the letter in order to avoid any potential failure and/or bed collapse and to return the" Retaining washers & Clevis Pins" after the replacement procedure is performed. If you have any questions, or need assistance in performing the replacement, call 330-965-9681.
- Initiated
- 2011-02-02
- Root cause
- Equipment maintenance
- Terminated
- 2014-05-14
- Distribution
- Nationwide distribution: USA including states of: AZ, IL, KY, MD, MI, OH, OR, PA, TX and VA.
- Recall number
- Z-0270-2013
- Recall status
- Terminated
- CDRH record ID
- 112493
- FDA classified
- 2012-11-08
- Recall event number
- 63083