Z-0270-2020: SCFE (slipped capital femoral epiphysis) Knobs for Driver, 6.5mm and 7.3 mm Orthopedic surgical instrume

Recall record Z-0270-2020 for Pega Medical Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0270-2020
Company in FDA record
Pega Medical Inc.
FDA record creation date
Nov 1, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for the screws to have become loosened during ultrasonic cleaning of the knobs and the screws may come out.

Source facts

Action
Pega Medical notified customers of the recall on about 10/02/2019, via "Urgent Field Safety Notice - Recall" letter. Customers were informed that SCFE Drivers, 6.5mm and 7.3 mm of the Free Gliding SCFE Screw System had issues with the knobs. Specifically the knobs have the potential to have loosened screws, or screws that have fallen out, as a result of the ultrasonic cleaning of the knobs. Customers were instructed to identify affected devices in inventory and return them to Pega Medical as soon as possible to address the issue. There returned devices can be returned via the FedEx information provided in the recall letter. Additionally customers were requested to provide the recall notification to all who need to be notified within your organization or any organization where the affected devices have been transferred. Contacts for questions include Mr. Enrique Garcia, at 450-688-5144 Ext 242, Email: egarcia@pegamedical.com and Mr. Kildare Molina, at 450-688-5144 Ext. 267, Email: kildare@pegamedical.com.
Initiated
2019-10-02
Root cause
Device Design
Terminated
2020-10-26
Distribution
Nationwide distribution to CA, FL, GA, HI, LA, MA, MO, NC, NY, OK. International distribution to AUSTRALIA, CANADA, DENMARK, GERMANY, ISRAEL, MEXICO, ROMANIA, RUSSIA, UNITED ARAB EMIRATES, UNITED KINGDOM
Recall number
Z-0270-2020
Recall status
Terminated
CDRH record ID
177004
FDA record created
2019-11-01
Recall event number
84017

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗