There is a potential for the patient to be exposed to cryogenic fluids (liquid air) dripping from the vent during a quench due to the location of the vent.
Source facts
Action
A Mandatory Service Letter, dated April 17, 2003, was sent to service technicians instructing the technicans to perform a mandatory safety modification.
Initiated
2003-04-17
Root cause
Other
Terminated
2008-10-29
Distribution
Throughout the United States and to Denmark, England, China, Italy, Mexico, Israel, and Puerto Rico.
Recall number
Z-0272-05
Recall status
Terminated
CDRH record ID
35828
FDA classified
2004-12-02
Recall event number
30416