Skip to content

Public FDA record

Z-0272-05: Intinion 1.5 T MRI Scanner.

Recall record Z-0272-05 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0272-05
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA classification date
Dec 2, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a potential for the patient to be exposed to cryogenic fluids (liquid air) dripping from the vent during a quench due to the location of the vent.

Source facts

Action
A Mandatory Service Letter, dated April 17, 2003, was sent to service technicians instructing the technicans to perform a mandatory safety modification.
Initiated
2003-04-17
Root cause
Other
Terminated
2008-10-29
Distribution
Throughout the United States and to Denmark, England, China, Italy, Mexico, Israel, and Puerto Rico.
Recall number
Z-0272-05
Recall status
Terminated
CDRH record ID
35828
FDA classified
2004-12-02
Recall event number
30416

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗