Open original FDA record ↗Recall record
- FDA identifier
- Z-0272-2016
- Company in FDA record
- Precision Medical, Inc.
- FDA classification date
- Nov 13, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
These lots of Oxygen Dial Regulators could have the settings label off by one position. When the indicator arrow points to OFF position, the regulator is actually on the half liter setting (0.50), for the 15 liter. For the 8 liter, instead of being in the OFF position, the regulator will be in the quarter liter setting (0.25).
Source facts
- Action
- The firm sent "Attention Field Corrective Action" letters, dated Oct 8, 2015, to customers. The letter identified the affected device and explained the reason for the recall. The letter provided instructions on how to check the regulator; and asked customers to call the number provided for a replacement if a regulator is labeled incorrectly.
- Initiated
- 2015-10-08
- Root cause
- Error in labeling
- Terminated
- 2016-10-27
- Distribution
- Distributed in the US to the states of PA, MA, MO, MT, ID, MI, MN, ND, KY, NY, NC, CA, AL, AZ, LA, OH, TX, GA, IN, and OK.
- Recall number
- Z-0272-2016
- Recall status
- Terminated
- CDRH record ID
- 141042
- FDA classified
- 2015-11-13
- Recall event number
- 72449