Open original FDA record ↗Recall record
- FDA identifier
- Z-0272-2022
- Company in FDA record
- Abbott Laboratories
- FDA record creation date
- Nov 19, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes.
Source facts
- Action
- On 09/29/2021, the Firm issued an "Urgent Medical Device Correction" Notification Letter via FedEx priority overnight express informing customers that the Firm has identified 3 potential performance issues with Software version 9.41 and earlier. The 3 issues are: 1. When an Error Code 3382, Unable to process test, internal wash pressure (x) error (y) pipettor occurs, the systems are incorrectly placed into a Scheduled Pause status rather than a Stopped status. As a result, the processing module continues to process tests after the hardware error is detected. This may cause incorrect results to be generated. 2. When configuring a Calibrator Sample Volume on the Configure Assays screen, if the user selects multiple assays, the calibrator sample volume from one assay may be carried into the calibrator sample volume for another assay. Incorrect calibrator sample volumes have the potential to generate incorrect calibration curves which may lead to incorrect results and delay of results due to the need to reconfigure assay parameters. 3. When performing a backup on the system while the iARM is replenishing the wash buffer at the same time, the iARM looses communication with the System Control Center. The loss of communication may cause the wash buffer container to overflow. This has the potential to lead to physical and chemical hazards. Customers are instructed to follow the Necessary Actions listed in Appendix A until software version 9.45 or software version 9.50 are installed. Appendix A states: For Issue (listed above): #1-The Necessary Actions to be taken by the customers until mandatory upgrade is completed - Stop the processing module and discard any results produced after an Error Code 3382. Perform Corrective Actions for Error Code 3382. For details, refer to System Operations Manual, Section 10: Troubleshooting and diagnostics. #2-The Necessary Actions to be taken by the customers until mandatory upgrade is completed - Verify the correct calibrator sample v
- Initiated
- 2021-09-29
- Root cause
- Software design
- Distribution
- U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. O.U.S.: AFGHANISTAN ALBANIA, ALGERIA, ANDORRA, ANGOLA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHAMAS, BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BELIZE, BENIN, BERMUDA, BOLIVIA, BOSNIA AND HERZEGOVINA, BOTSWANA, BRAZIL, BRUNEI, BULGARIA, BURKINA FASO, BURUNDI, CAMBODIA, CAMEROON, CANADA, CAYMAN ISLANDS, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DEMOCRATIC REP. OF CONGO, DENMARK, FAROE ISLANDS, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ERITREA, ESTONIA, ETHIOPIA, FINLAND, FRANCE, GEORGIA, GERMANY, GHANA, GREECE, GREENLAND, GUATEMALA, GUINEA, HAITI, HONDURAS, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, IVORY COAST, JAMAICA, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, KYRGYZSTAN, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MADAGASCAR, MALAWI, MALAYSIA, MALDIVES, MALI, MALTA, MAURITANIA, MAURITIUS, MEXICO, MOLDOVA, MONTENEGRO, MONTSERRAT, MOROCCO, MOZAMBIQUE, MYANMAR, NAMIBIA, NEPAL, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, NICARAGUA, NIGERIA, NORWAY, OMAN, PAKISTAN, PANAMA, PARAGUAY, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, ROMANIA, RUSSIA, RWANDA, SAUDI ARABIA, SENEGAL, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, ST. LUCIA, ST. VINCENT, SURINAME, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TAJIKISTAN, TANZANIA, THAILAND, TRINIDAD & TOBAGO, TUNISIA, TURK & CAICOS, TURKEY, UAE, UGANDA, UK, UKRAINE, URUGUAY, UZBEKISTAN, VIETNAM, WEST BANK/GAZA STRIP, YEMEN, ZAMBIA, ZIMBABWE,
- Recall number
- Z-0272-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189936
- FDA record created
- 2021-11-19
- Recall event number
- 88880