Software defect causes patient identification number to be truncated.
Source facts
Action
Firm issued a letter notifying users of systems that when transferring image data to a DICOM server for any patient born on January 1, to verify that the Patient ID is correct on the DICOM server after the transfer. Firm will modify the software by January 2005.
Initiated
2004-08-24
Root cause
Other
Terminated
2005-09-02
Distribution
Nationwide.
Recall number
Z-0273-05
Recall status
Terminated
CDRH record ID
35832
FDA classified
2004-12-02
Recall event number
30420