Skip to content

Public FDA record

Z-0273-2010: Bard Magnum Disposable Biopsy Core Needle 16 gauge, Catalog Number: MN1610. Bard Magnum Disposable Biops

Recall record Z-0273-2010 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0273-2010
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Nov 17, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated after the firm determined that some Magnum Disposable Biopsy Core Needle 16 gauge lots may exhibit an exposed sample chamber, due to extended stylet length resulting from a loose hub connection. The exposed sample chamber will typically result in an inability to obtain the biopsy sample. In the majority of reported cases, there was no injury to the patient or user. When us

Source facts

Action
The recall notification was initiated by the firm issuing an urgent medical device notification to it's consignees on October 08, 2009 via USPS requesting them to check all inventory locations for the affected products, remove and return affected products to Bard. The consignees were instructed to complete and return a recall effectiveness check form. Bard provided this 800 number 1-800-321-4254 for any questions consignees may have.
Initiated
2009-10-08
Root cause
Process design
Terminated
2012-04-02
Distribution
Nationwide and internationally to Canada Mexico Algeria Andorra Austria Bahrain Belarus Belgium Croatia Cyprus Czech Republic Denmark Finland France French Guiana Germany Gibraltar Greece Hungary Ireland Israel Italy Ivory Coast Kenya Kuwait Latvia Lebanon Luxembourg Malta Martinique Mauritius Morocco Netherlands New Caledonia Nigeria Norway Oman Poland Portugal Republic of Kosovo Republic of Macedonia Reunion Romania Russia Saudi Arabia Senegal Serbia Seychelles Slovenia South Africa Spain Sweden Switzerland Syria Tunisia Turkey Ukraine United Arab Emirates United Kingdom Australia New Zealand Canada Japan Bangladesh China Hong Kong India Indonesia Korea Malaysia Pakistan Singapore Sri Lanka Taiwan Thailand Vietnam Argentina Bahamas Barbados Brazil Chile Colombia Dominican Republic Mexico Peru and Suriname
Recall number
Z-0273-2010
Recall status
Terminated
CDRH record ID
86079
FDA classified
2009-11-17
Recall event number
53615

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗