Z-0273-2022: Cardioblate CryoFlex Surgical Ablation Console

Recall record Z-0273-2022 for Medtronic Perfusion Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0273-2022
Company in FDA record
Medtronic Perfusion Systems
FDA record creation date
Nov 19, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential for intermittent electrical connectivity between the console and probe.

Source facts

Action
In the United States (US), beginning 27-SEP-2021 Medtronic initiated communications of an Urgent Medical Device Correction letter via 2-day UPS delivery to consignees who have received affected product according to Medtronic records. Outside of the United States (OUS), beginning 27-SEP-2021, a written Urgent Medical Device Correction letter is being delivered via regionally approved methods to consignees who have received affected product according to Medtronic records. Consignees are asked to review the Medtronic Urgent Medical Device Correction notification and follow the Recommended Actions, including quarantining affected Cardioblate CryoFlex Surgical Ablation Consoles in their inventory for servicing.
Initiated
2021-09-27
Root cause
Process control
Distribution
Distribution in US - IL, KS, and UT Distribution outside of US to Australia and South Korea
Recall number
Z-0273-2022
Recall status
Open, Classified
CDRH record ID
189719
FDA record created
2021-11-19
Recall event number
88845

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗