Potential weld defect in the Overhead Tube Support assembly which connects the tube assembly to the telescoping arm. This could cause the tube assembly to fall, injuring anyone near the unit.
Source facts
Action
Letters dated 10/20/2005 were hand carried by Kodak Field Service Reps. Kodak Field Engineers inspected units and found the two defective units (both in China).
Initiated
2005-10-20
Root cause
Other
Terminated
2005-12-30
Distribution
DOMESTIC: 12 units shipped to 12 medical facilities in FL, KY, MI, NY, OH, PA, and SC. FOREIGN: 12 units shipped to 7 hospitals in China, 1 warehouse in China; 1in the possession of Chinese Customs (government); and 1 is in a warehouse in Russia.
Recall number
Z-0274-06
Recall status
Terminated
CDRH record ID
42757
FDA classified
2005-12-08
Recall event number
33931