Z-0275-06: Baxter Meridian Hemodialysis Instrument, product codes 5M5576 and 5M5576R; Baxter Healthcare Corporation,

Recall record Z-0275-06 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0275-06
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Dec 14, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Inadequate instructions for routing single tubing lines through the double-tubing clips on the front of the instrument to eliminate the risk of kinks in the tubing, which have been associated with a death and a serious injury.

Source facts

Action
Urgent Device Recall letters dated 9/28/05 were sent to all Meridian accounts on the same date, to the attention of the Hemodialysis Administrator, with copies sent to the Chief Technician and Hemodialysis Charge Nurse. The letters informed the accounts of teh death and serious injury related to kinks in blood tubing sets used on the Meridian machine, and that there is an increased risk of kinks to occur at the double-tubing clips mounted on the front of the Meridian. The risk increases when a single tubing line is routed through both retainers on either clip. The letter included diagrams showing the correct and incorrect use of the clips, provided labels, with placement instructions, to be placed on the Meridian machine indicating the correct and incorrect placement of tubing, and informed the users that replacement clips with single channels will be installed when they become available. Follow-up letters were sent to the accounts on 12/22/05, providing training material to ensure that the accounts are aware of proper blood tubing set-up and understand the conditions which may result in hemolysis, and on 12/23/05, informing them that the replacement clips were available for installation, giving the accounts the option of having Baxter install them or sending the clips to the account for self-installation.
Initiated
2005-09-28
Root cause
Other
Terminated
2006-11-27
Distribution
Nationwide and internationally to Mexico and Asia.
Recall number
Z-0275-06
Recall status
Terminated
CDRH record ID
42188
FDA classified
2005-12-14
Recall event number
33632

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗