Skip to content

Public FDA record

Z-0275-2016: GMK Hinge knee prosthesis; Model Number: 02.09.TA410; Orthopedics and Orthopedic Surgery

Recall record Z-0275-2016 for Medacta Usa. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0275-2016
Company in FDA record
Medacta Usa
FDA classification date
Nov 14, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect parts: Packages of the GMK Hinge Tibial Augmentation size 4-10 mm have 5 mm screws instead of 10 mm screws.

Source facts

Action
On 4/16/15 all distributors with recalled product were immediately notified of the recall by phone. On 4/20/15, all distributors were further notified of the recall via email. The URGENT MEDICAL DEVICE RECALL letter dated 4/17/2015 the firm described the initiation of a recall for regarding the GMK HINGE Tibial Augmentation Size 4-10mm Reference 02.09TA410 / Lot 145129. The firm provided the following recall actions to be performed: 1.Check your inventory for the device listed above. If you find it, IMMEDIATELY place it in quarantine and DO NOT use it in surgery. 2.If you have further distributed this device to other customers please notify me and initiate their return to Medacta USA. 3.We will provide you with an update on a replacement within the next few days. The firm also provided a contact for customers requiring additional information. Mike Loiterman Director of Regulatory, Quality and Compliance Medacta USA Phone: 312548-9971Fax: 312-546-6881MLoiterman@medacta.us.com
Initiated
2015-04-17
Root cause
Employee error
Terminated
2017-01-05
Distribution
Distributed in the states of IL, OH and UT, and the countries of in Austria, Spain, Australia, France and Switzerland.
Recall number
Z-0275-2016
Recall status
Terminated
CDRH record ID
136166
FDA classified
2015-11-14
Recall event number
71121

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗