Z-0276-05: OTC Pregnancy Tests. Name: Sure Check hCG Midstream Pregnancy Test --- Description: the Sure Check hCG Mi

Recall record Z-0276-05 for Chembio Diagnostic System Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0276-05
Company in FDA record
Chembio Diagnostic System Inc
FDA classification date
Dec 3, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Poor sealing of pouches and degradation of products. Firm received an e-mail from a distributor of an increase in complaints. In response to the complaints, the distributor conducted its own functional testing and noticed a degradation in performance of the two lots. Chembio's investigation revealed invalid results (no control lines) and some false negatives.

Source facts

Action
Recall letters on 10/6/04, followed by a second letter on/about 10/20/04. Recall letters mailed on 11/9/04 to cover recall expansion for 4 additional lots.
Initiated
2004-10-06
Root cause
Other
Terminated
2005-03-24
Distribution
Distributors in KS, TN, NC, MA, NY, KY, FL, NV, and Puerto Rico for further distribution to their retail customers. Products also distributed to accounts in Barbados WI, Israel, Costa Rica and Canada.
Recall number
Z-0276-05
Recall status
Terminated
CDRH record ID
35616
FDA classified
2004-12-03
Recall event number
30288

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗