Z-0276-2022: HemosIL ReadiPlasTin (20 mL), Part number 0020301400. An in vitro diagnostic thromboplastin reagent.

Recall record Z-0276-2022 for Instrumentation Laboratory. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0276-2022
Company in FDA record
Instrumentation Laboratory
FDA classification date
Nov 23, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.

Source facts

Action
On October 13, 2021, the firm sent customers Urgent Medical Device Removal letters via email. These emails were followed by certified letters sent starting the same day. ***On February 17, 2022, the firm expanded the recall to include an additional lot of product.*** The customers were provided with instructions to run quality controls with each vial. Customers should discard any vials with failed quality controls. For technical questions, please contact the IL Technical Support Center at 1-800-678-0710, Option # 2 (ACL Coagulation Products). For alternate lot availability questions, please contact your IL Customer Service Representative at 1-800-955-9525, Option # 0.
Initiated
2021-10-13
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, IA, IL, KS, ME, MI, NC, NY, OK, SC, and VA. The countries of Canada, Chile, China, Costa Rica, Italy, Japan, and South Korea. ***Updated 3/3/22*** Expansion includes states of IN, MN, and NC.
Recall number
Z-0276-2022
Recall status
Open, Classified
CDRH record ID
190052
FDA classified
2021-11-23
Recall event number
88930

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗