Poor sealing of pouches and degradation of products. Firm received an e-mail from a distributor of an increase in complaints. In response to the complaints, the distributor conducted its own functional testing and noticed a degradation in performance of the two lots. Chembio's investigation revealed invalid results (no control lines) and some false negatives.
Source facts
Action
Recall letters on 10/6/04, followed by a second letter on/about 10/20/04. Recall letters mailed on 11/9/04 to cover recall expansion for 4 additional lots.
Initiated
2004-10-06
Root cause
Other
Terminated
2005-03-24
Distribution
Distributors in KS, TN, NC, MA, NY, KY, FL, NV, and Puerto Rico for further distribution to their retail customers. Products also distributed to accounts in Barbados WI, Israel, Costa Rica and Canada.
Recall number
Z-0277-05
Recall status
Terminated
CDRH record ID
35617
FDA classified
2004-12-03
Recall event number
30288