Open original FDA record ↗Recall record
- FDA identifier
- Z-0277-2023
- Company in FDA record
- OrthoPediatrics Corp
- FDA classification date
- Nov 22, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Bolt breakage on the head or threaded portion, which may cause loss of strut continuity with frame. This may cause loss of deformity correction and/or disruption of bone regeneration.
Source facts
- Action
- On September 12, 2022, the firm notified distributors of the recall. Distributors were asked to notify and retrieve affected devices from end users. Affected product must be returned to OrthoPediatrics. Replacement product will be provided. Customers should contact customer service to arrange return of the affected product. Customers who have patients in frames should be made aware of the possibility of Bolt breakage and continue monitoring their patients. Any questions regarding this recall can be directed to Joel Batts, Senior VP of Clinical and Regulatory Affairs, at OrthoPediatrics at jbatts@orthopediatrics.com or 574-367-5396.
- Initiated
- 2022-09-12
- Root cause
- Under Investigation by firm
- Distribution
- Domestic distribution to Foreign distribution to
- Recall number
- Z-0277-2023
- Recall status
- Open, Classified
- CDRH record ID
- 196726
- FDA classified
- 2022-11-22
- Recall event number
- 91098