Philips MRx devices have an EtCO2 module that may contain contamination; this contamination (will/may) cause the device to be unable to display the EtCO2 or CO2 waveform (capnogram).
Source facts
Action
The Philips Urgent Medical Device Correction notification/ Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. The correction will consist of providing a hardware correction, free of charge, for all units affected by this issue.
Initiated
2017-09-28
Root cause
Nonconforming Material/Component
Terminated
2018-10-12
Distribution
20 units affected. 18 domestic, 2 foreign.
Recall number
Z-0278-2018
Recall status
Terminated
CDRH record ID
160142
FDA record created
2017-12-21
Recall event number
78698