Open original FDA record ↗Recall record
- FDA identifier
- Z-0279-2012
- Company in FDA record
- Steris Corporation
- FDA classification date
- Nov 23, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Some hydraulic column cylinders installed in certain 5085 and 5085SRT Surgical Tables were assembled by their supplier with incorrect snap rings and an anomaly was present in the tilt cylinder. This error may affect the user's ability to move the table top out of the full right tilt position.
Source facts
- Action
- On October 17, 2011, Steris sent 'Urgent Voluntary Field Correction Notices' to their customers via Fedex with tracking numbers for delivery confirmation. The letter identifies the affected product and notifies the user/customers of the problem with the surgical tables. The notification letter goes on to inform the customers that a STERIS Service Representative will be contacting them to arrange for the field correction of the affected unit(s). This involves the replacement of the hydraulic column cylinder of each surgical table(s).
- Initiated
- 2011-10-17
- Root cause
- Nonconforming Material/Component
- Terminated
- 2013-03-19
- Distribution
- Nationwide distribution.
- Recall number
- Z-0279-2012
- Recall status
- Terminated
- CDRH record ID
- 104997
- FDA classified
- 2011-11-23
- Recall event number
- 60305