Based on a retrospective review of quality data, AC Control boards used in the manufacture of the
HX2 Temperature Management System were identified as being manufactured using the incorrect
flux material. Use of the incorrect flux may contribute to premature board failure. A search of
TCVS complaints revealed 3 complaints received related to the HX2 failure. Each failure occurred
during set up
Source facts
Action
TERUMO sent an Urgent Medical Device Recall letter dated October 15, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Terumo has determined that a field correction is required to notify affected customers that the boards may fail, instruct the users that in a case of failure to swich to the second channel on the device or use a back-up unit, and that TCVS will replace the affected boards. Customers with questions were instructed to call 1-800-621-2818. For questions regarding this recall call 734-741-6173.
Initiated
2012-10-16
Root cause
Nonconforming Material/Component
Terminated
2013-07-11
Distribution
Nationwide Distribution including CA, HI, KS, LA, MI, MO, OR, PA, UT, and WI
Recall number
Z-0279-2013
Recall status
Terminated
CDRH record ID
113535
FDA classified
2012-11-13
Recall event number
63344