Skip to content

Public FDA record

Z-0279-2016: MiniArc Pro Single incision Sling System, Model number / Product Reference 9000261, Lot 936405xxx. Desig

Recall record Z-0279-2016 for Astora. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0279-2016
Company in FDA record
Astora
FDA classification date
Nov 16, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The lot number on the outer carton label did not match the lot number printed on the tray label or identification labels inside the box. The identification labels inside the box are placed on medical records and other documents associated with the procedure. Use of incorrectly identified labels may prevent direct traceability of the implanted unit in the future.

Source facts

Action
Consignees were hand delivered by sales representatives a Astora Women's Health "Urgent Recall Notice". The letter was addressed to Physician, Health Care Professional, Nurse & Risk Manager. The letter described the problem and the product involved in the recall. Consignees were advised to check their inventory and return all affected product, notify those that might be affected and to complete and return the Acknowledgement Form. For questions contact ASTORA Women's Health Customer Service at 1-844-879-0787.
Initiated
2015-10-26
Root cause
Labeling mix-ups
Terminated
2016-03-09
Distribution
Distributed in the states of AZ, CA, DE, IL, KS, NE, NJ, NY, OH, PA, TX
Recall number
Z-0279-2016
Recall status
Terminated
CDRH record ID
141509
FDA classified
2015-11-16
Recall event number
72552

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗