Open original FDA record ↗Recall record
- FDA identifier
- Z-0279-2020
- Company in FDA record
- Baxter Healthcare Corporation
- FDA record creation date
- Nov 4, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Communication error alarms may result in interruption of therapy, delay in therapy, or blood loss due to non-restitution of blood in the extracorporeal circuit.
Source facts
- Action
- Baxter Healthcare notified customers on about 10/04/2019, via "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were informed that Prismaflex Control Units with software versions 7.20 and below have the potential to produce communication errors. Instructions included that operators may continue to use the control units until the upgrade is performed, but if a communication error alarm does occur, follow the instructions presented on the graphical user interface and/or in the operators manual. It must be reinforced that it is important to manually return the extracorporeal blood to the patient, and re-train users on this process, outlined on page 10:57 of the operators manual, if necessary. A local Baxter service representative will contact your facility to determine the correction plan and schedule the software upgrade. If the affected devices was purchased from Baxter, complete the provided Baxter Customer Reply Form and return it to Baxter by faxing it to 224-270-5457 or scanning and e-mailing it to fca@baxter.com. If you distribute this product to other facilities or departments within your institution, please forward a copy of this communication to them. For a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures. Questions regarding this recall can be directed to Baxter Corporate Product Surveillance at 800-437-5176, between the hours of 8:00 am and 5:00 pm Central Time, Monday through Friday.
- Initiated
- 2019-10-04
- Root cause
- Software design
- Distribution
- Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY. International distribution to Australia, Bangladesh, Brunei, China, Hongkong, Indonesia, India, Japan, New Zealand, Taiwan, Korea, Singapore, Malaysia, Macau, Thailand, Vietnam, Canada, Argentina, Costa Rica, Cuba, Dominican Republic, Haiti, Barbados, Venezuela, St. Kitts, and EMEA.
- Recall number
- Z-0279-2020
- Recall status
- Open, Classified
- CDRH record ID
- 176501
- FDA record created
- 2019-11-04
- Recall event number
- 83852