Open original FDA record ↗Recall record
- FDA identifier
- Z-0279-2025
- Company in FDA record
- Accriva Diagnostics, Inc.
- FDA classification date
- Oct 31, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Values used to calibrate whole blood oximeter instruments were observed to be higher than normal, which may result in incorrect patient results where oxygen levels are higher than actual results, which may lead to patient care mismanagement requiring medical intervention.
Source facts
- Action
- On 9/27/24, a recall notice was sent to a customer informing them to do the following: 1) This recall notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where the potentially affected device has been transferred.. 2. Call Technical Support 1-800-678-0710 to initiate a Service Call and Returned Material Authorization 3. Complete and return the customer response form via email to For technical questions contact Technical Support at the number above or email techsupport.na@werfen.com.
- Initiated
- 2024-09-27
- Root cause
- Process change control
- Distribution
- US: AZ
- Recall number
- Z-0279-2025
- Recall status
- Completed
- CDRH record ID
- 210408
- FDA classified
- 2024-10-31
- Recall event number
- 95497