- Recall classification
- Class II
Reason stated in the source
The packaging of a sterile battery, intended to power a surgical screwdriver, may contain Tyvek shavings inside the sterile package.
Source facts
- Action
- On 9/23/24, an "URGENT" recall notice was emailed to the customer informing of the product, issue, risk and actions to be taken. Customers were instructed to do the following: 1) Please immediately cease distribution and use of these lots. Inventory and distribution records should be evaluated to ensure that affected product is removed from any potential point of use. 2) Product should be returned to the recalling firm. 3) The Recall Acknowledgement and Return Form should be completed and returned via email to Recall@Pro-Dex.com or FAX 949.769.3281. Questions or concerns can be directed to the firm at 949-769-3200 or Recall@Pro-Dex.com
- Initiated
- 2024-09-23
- Root cause
- Component design/selection
- Distribution
- US distribution to FL only.
- Recall number
- Z-0280-2025
- Recall status
- Open, Classified
- CDRH record ID
- 210457
- FDA classified
- 2024-10-31
- Recall event number
- 95510