The brake/steer pedal may seize, resulting in the brakes not working.
Source facts
Action
Consignees were notified by letter dated August 10, 2009 that Stryker will contact them to upgrade the stretchers within 6 months. The recall was expanded in scope to include additional serial numbers by a second letter dated September 17, 2009, which was issued to additional consignees who had not been previously notified of the recall, as well as the consignees who had been issued the August 10, 2009 letter.
Initiated
2009-08-10
Root cause
Nonconforming Material/Component
Terminated
2010-10-29
Distribution
Nationwide, Argentina, Australia, Brazil, Canada, Chile, China, France, Germany, Greece, India, Italy, Japan, Korea, Latin America, Malaysia, Mexico, Netherlands, Philippines, Poland, Portugal, Singapore, South Africa, Spain, Switzerland, Taiwan, and United Kingdom,
Recall number
Z-0281-2010
Recall status
Terminated
CDRH record ID
82581
FDA classified
2009-11-17
Recall event number
52229
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.