Ray Search Laboratories found an issue with the dose calculation for Elekta motorized wedges in Ray Station 3.0, 3.5 and 4.0. The error presents when a wedge is used for a beam and the wedge angle differs from 60 degrees.
Source facts
Action
Medical Device Correction Letters dated 10/23/13 were e-mailed to the user accounts on 10/23/13.
Initiated
2013-10-23
Root cause
Software design
Terminated
2017-02-17
Distribution
US Distribution in the states of: California, Colorado, Delaware, Florida, Maryland, Missouri, New Jersey, New York, Ohio, Pennsylvania, Tennessee, and Virginia.
Recall number
Z-0281-2014
Recall status
Terminated
CDRH record ID
122696
FDA classified
2013-11-14
Recall event number
66600