The power supply unit for the printer used with the device can potentially overheat or cause a fire hazard because it could contain an improperly-mixed phosphorus compound.
Source facts
Action
The recalling firm issued letters dated 9/3/2018 via email on 9/17/2018. A second letter dated 9/24/2018 was issued to update the acknowledgment form to include the number of affected power supply units the consignee destroyed.
Initiated
2018-09-17
Root cause
Nonconforming Material/Component
Terminated
2020-08-21
Distribution
Distribution was nationwide. There was also government/military distribution. Foreign distribution was made to Canada.
Recall number
Z-0281-2019
Recall status
Terminated
CDRH record ID
167333
FDA record created
2018-10-26
Recall event number
81036