Open original FDA record ↗Recall record
- FDA identifier
- Z-0281-2020
- Company in FDA record
- GE Healthcare, LLC
- FDA record creation date
- Nov 4, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Fasteners potentially torqued to a value less than the specified value and can potentially cause the canopy to become detached and fall.
Source facts
- Action
- GE Healthcare notified customers on about 07/26/2019 via phone calls and a "URGENT MEDICAL DEVICE CORRECTION" letter. Customers were instructed to immediately stop use of and segregate the affected devices, that GE Healthcare would correct all affected products at no cost, and a GE Healthcare representative will contact them to arrange for the correction. Questions or concerns can be directed to GE Healthcare Service at 1-800-437-1171 or your local Service Representative. A second customer letter was sent on 08/05/2019 to all customer. The second letter contained the same instructions, but provided an updated list of affected serial numbers.
- Initiated
- 2019-07-26
- Root cause
- Nonconforming Material/Component
- Distribution
- Worldwide distributions - US Nationwide in the states of CA, CT, FL, ID, IL, KS, MI, NC, NC, NJ, NM, PA, TX, VA. Countries of China, France, Germany, Korea, Republic of China.
- Recall number
- Z-0281-2020
- Recall status
- Open, Classified
- CDRH record ID
- 175629
- FDA record created
- 2019-11-04
- Recall event number
- 83532