Open original FDA record ↗Recall record
- FDA identifier
- Z-0282-2012
- Company in FDA record
- Varian Medical Systems, Inc.
- FDA classification date
- Nov 23, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Instructions For Use ( IFU) distributed with product prior to 2004 lacks the correct limit to the sterilization cycles and checks regarding the firm seating of the blocking washer.
Source facts
- Action
- Varian Medical Systems sent a Medical Device Recall (PNL-GM-42697-2) in May 2010 for the blocking washer on the flexible probe, Varian part number GM11002420, a component of the Segmented Cylinder Applicator Set. The letter was to inform users that a flexible probe with an unstable or shifted blocking washer must not be used due to the potential for treatment length error. The blocking washer may move from rough handling or sterilization. Users were told to discard any instances of the obsolete Instructions for Use (IFU) in use and replace it with the IFU (PGM11004150 4 March 24, 2010.pdf). A Receipt Verification Card was included requesting consignees to fill out, fax or e-mail immediately to Varian BrachyTherapy.
- Initiated
- 2010-05-20
- Root cause
- Labeling Change Control
- Terminated
- 2011-12-05
- Distribution
- Worldwide Distribution: USA, Germany, Belgium, Canada, Mexico City, Spain, Brazil, Switzerland, Uzbekistan, India
- Recall number
- Z-0282-2012
- Recall status
- Terminated
- CDRH record ID
- 103990
- FDA classified
- 2011-11-23
- Recall event number
- 59913