The assigned value for Rheumatoid Factor over recovers against the reference material RF serum, 1st British standard from NIBSC. Ref = 64/002. The calibrator values have been adjusted to re-align to the reference material.
Source facts
Action
1. Review your reagent inventory and replace the value sheet with the new values provided.
2. Review results generated with the affected batches in line with the clinical profile of the patient.
3. Discuss the contents of this notice with your Medical Director.
4. Complete and return the response form 12187-QA to teclhnical.services@randox.com within five working days.
Initiated
2019-08-09
Root cause
Nonconforming Material/Component
Terminated
2021-01-08
Distribution
US Nationwide distribution in the states of WV, ME, NJ, CA, and Puerto Rico.
Recall number
Z-0282-2020
Recall status
Terminated
CDRH record ID
176361
FDA record created
2019-11-04
Recall event number
83791