Z-0282-2022: Medtronic CareLink SmartSync Device Manager application software (D00U005) used by the following Medtroni

Recall record Z-0282-2022 for Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0282-2022
Company in FDA record
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Nov 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The processing of collected episode data may fail due to a software error.

Source facts

Action
The recalling firm began issuing the correction notification dated October 2021 on 10/12/2021 via 2-day UPS delivery that was addressed to risk managers or clinical managers. The letter recommends physicians follow normal clinical practices given the implanted Cobalt/Crome devices will continue to operate as programmed. If a failure to interrogate a Cobalt or Crome device occurs with a SmartSync programmer, the physician is to confirm that the SmartSync application software has been updated to D00U005 version 5.0.0 (or higher). They are to contact their Medtronic representative or Tachy Technical Services at 800-723-4636 for assistance with retrieving the session data. If a CareLink transmission is attempted, but the transmission is not viewable on the CareLink network (i.e., the transmission is missing from the transmission list for the patient), the physician is to contact Medtronic Technical Services at 800-723-4636 for assistance. The notice is to be shared with those who need to be aware within the consignee's organization or with any organization where the SmartSync programmers may have been transferred. Consignees are asked to sign and return the enclosed confirmation form and confirm that the SmartSync tablets have been updated to Cobalt/Crome application software D00U005, version 5.0.0 or higher.
Initiated
2021-10-12
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide. There was military/government distribution. The countries of Andorra, Australia, Austria, Belgium, Brunei Darussalam, Canada, Czech Republic, Denmark, Finland, France, French Guiana, French Polynesia, Germany, Greece, Guadeloupe, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Liechtenstein, Luxembourg, Martinique, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Puerto Rico, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
Recall number
Z-0282-2022
Recall status
Open, Classified
CDRH record ID
189977
FDA record created
2021-11-24
Recall event number
88907

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗