- Action
- The recalling firm began issuing the correction notification dated October 2021 on 10/12/2021 via 2-day UPS delivery that was addressed to risk managers or clinical managers. The letter recommends physicians follow normal clinical practices given the implanted Cobalt/Crome devices will continue to operate as programmed. If a failure to interrogate a Cobalt or Crome device occurs with a SmartSync programmer, the physician is to confirm that the SmartSync application software has been updated to D00U005 version 5.0.0 (or higher). They are to contact their Medtronic representative or Tachy Technical Services at 800-723-4636 for assistance with retrieving the session data. If a CareLink transmission is attempted, but the transmission is not viewable on the CareLink network
(i.e., the transmission is missing from the transmission list for the patient), the physician is to contact Medtronic Technical Services at 800-723-4636 for assistance.
The notice is to be shared with those who need to be aware within the consignee's organization or with any organization where the SmartSync programmers may have been transferred.
Consignees are asked to sign and return the enclosed confirmation form and confirm that the SmartSync tablets have been updated to Cobalt/Crome application software
D00U005, version 5.0.0 or higher.
- Initiated
- 2021-10-12
- Root cause
- Software design
- Distribution
- Worldwide distribution - US Nationwide. There was military/government distribution. The countries of Andorra, Australia, Austria, Belgium, Brunei Darussalam, Canada, Czech Republic, Denmark, Finland, France, French Guiana, French Polynesia, Germany, Greece, Guadeloupe, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait,
Liechtenstein, Luxembourg, Martinique, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Puerto Rico, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.
- Recall number
- Z-0282-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189977
- FDA record created
- 2021-11-24
- Recall event number
- 88907