The Spectra System/Legacy Guided Surgery Handle Kit had guided handle inserts that were out of Implant Direct specifications, from the results of complaints submitted by clinicians regarding the 3.4mm guide insert that will not allow the 3.4 drill to pass through, since the diameter is too small. In addition, the 2.3mm guide was too large.
Source facts
Action
On November 22, 2017, the recalling firm will send consignees customer letters with replacement product, requesting nonconforming product to be returned to Implant Direct with return labels and packaging. The recalling firm will then initiate calls 24 hours after the communication is sent to the clinician; initiate second calls 72 hours after the communication is sent; and initiate third calls 5 working days after the communication is sent. For further questions, please call (818) 444-3300.
Initiated
2017-11-15
Root cause
Labeling mix-ups
Terminated
2018-10-09
Distribution
Worldwide Distribution - USA (nationwide) Distribution and to the countries of : Europe, Canada, Australia, Egypt, Jordan, and Lebanon.
Recall number
Z-0283-2018
Recall status
Terminated
CDRH record ID
159820
FDA record created
2017-12-21
Recall event number
78583