One da Vinci Xi Endoscope Controller in the field was improperly calibrated during servicing. The result of this improper calibration may result in the tip of the endoscope may reach higher temperature causing the potential for thermal tissue injury and/or overly bright images.
Source facts
Action
Send formal customer letter communication to the affected site. The firm reported that a representative scheduled an on-site visit to replace the affected unit.
Initiated
2019-06-17
Root cause
Equipment maintenance
Terminated
2022-07-18
Distribution
US Nationwide distribution including in the state of Washington.
Recall number
Z-0283-2021
Recall status
Terminated
CDRH record ID
179714
FDA record created
2020-10-23
Recall event number
84963