Skip to content

Public FDA record

Z-0283-2024: Philips EPIQ Diagnostic Ultrasound System Models: EPIQ 7G(model # 795200), EPIQ 7C (Model # 795201), EPI

Recall record Z-0283-2024 for Philips Ultrasound, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0283-2024
Company in FDA record
Philips Ultrasound, Inc.
FDA classification date
Nov 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is the potential that diagnostic ultrasound systems operating with software version 10.0 may present a delayed image instead of a real time image.

Source facts

Action
On October 9, 2023, Philips issued a Urgent Medical Device Correction notification to affected consignees. Philips asked consignees to take the following actions: 1. Identify if your firm has an affected product. 2. In the 2D tab in the system touchscreen, disable the AutoSCAN feature every time a transducer or a Transducer preset is selected to prevent the system from potentially experiencing image lag issue. The AutoSCAN feature is disabled when the button is not highlighted in amber. 3. Please complete and return the attached response form to Philips promptly and no later than 15 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter, understanding of the issue, and required actions to be taken. 4. Circulate this Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue. Please retain this letter with your system(s) and ensure the letter is in a place likely to be seen/viewed. 5. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, or by regular mail, or by fax.
Initiated
2023-10-09
Root cause
Software design
Distribution
US: AZ, CA, CT, FL, GA, ID, IL, KY, MD, MT, NE, NV, OH, TN, TX, UT, VA OUS: Canada
Recall number
Z-0283-2024
Recall status
Completed
CDRH record ID
203691
FDA classified
2023-11-09
Recall event number
93218

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗