Open original FDA record ↗Recall record
- FDA identifier
- Z-0284-2012
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Nov 23, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Errors could potentially impact image interpretation. The acquisition software is exhibiting a defect in the energy window position when an isotope is peaked. There is a potential patient safety issue when running syngo MI Applications version VA60A with Foresight detectors firmware version v1.5.9.12 on the Symbia systems.
Source facts
- Action
- Siemens Medical Solutions USA, Inc. sent an "Urgent Field Correction Recall" letter dated October 14, 2011 to all affected customers. The letters informed customers of the acquisition software defect in the energy window position when an isotope is peaked, which could potentially impact image interpretation, and that the AEB firmware version on their imaging system has been changed to a previously released version until acquisition software service pack is available. For questions or assistance they were instructed to contact Siemens Medical Solutions USA, Inc. at (800) 888-7436.
- Initiated
- 2011-10-14
- Root cause
- Component change control
- Terminated
- 2012-10-16
- Distribution
- Florida, Illinois and Texas
- Recall number
- Z-0284-2012
- Recall status
- Terminated
- CDRH record ID
- 104522
- FDA classified
- 2011-11-23
- Recall event number
- 60142