Skip to content

Public FDA record

Z-0284-2027: ETHILON Nylon Suture, Product Code 697G

Recall record Z-0284-2027 for Ethicon, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0284-2027
Company in FDA record
Ethicon, LLC
FDA classification date
Oct 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Nylon suture indicated for use in general soft tissue approximation and/or ligation may have needles with distorted curvature, that may break during use, or could cause additional soft tissue or vessel injury, including tearing or bleeding intraoperatively.

Source facts

Action
On 9/3/2026, field safety notices were mailed to customers who were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). If you have product subject to this recall, please maintain a copy of this notice with the quarantined product and keep a copy for your records. 2. Return affected product to firm. 3. Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return and forward them this recall notice. 4. Complete and return the Business Reply Form (BRF) via email to RA-ITSUS-JJMCanada@ITS.JNJ.com
Initiated
2026-09-23
Root cause
Process control
Distribution
International distribution to the countries of JP, CA, AU.
Recall number
Z-0284-2027
Recall status
Open, Classified
CDRH record ID
222959
FDA classified
2026-10-23
Recall event number
99947

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗