Z-0285-2019: Magellan LeadCare II Blood Lead Test Kit. In Vitro Test for the measurement of lead in fresh whole blood.

Recall record Z-0285-2019 for Magellan Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0285-2019
Company in FDA record
Magellan Diagnostics, Inc.
FDA record creation date
Oct 26, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Defect with the calibration button for lot 1808M. Level 1 and Level 2 Quality Control (QC) values will generate results of HIGH (>65 g/dL). Results for patient testing with this lot will be displayed as HIGH.

Source facts

Action
The firm, Magellan, notified the Customers (end users) by telephone or email on August 8, 2018 to provide notification of the deficiency of the LeadCare II Blood Lead Test Kit having Lot Number 1808M. Replacement Buttons and an "URGENT MEDICAL DEVICE CORRECTION" letter dated August 9 2018 ( were shipped on August 9, 2018, for delivery on Friday, August 10, 2018.) The letter described the product, problem and actions to be taken. The customers were instructed to do the following: 1. Do NOT use the Yellow calibration button marked 1808M. 2. Discard the Yellow calibration button that came with the LeadCare II Test Kit Lot 1808M. 3. Place the new Blue calibration button with the test kit and use for controls and patient testing with Lot 1808M. 4. Inform your staff of the calibration button exchange. 5. Complete the fax form record included with this letter and fax to 978-600-1480 or email a scanned copy of the completed form to reply@magellandx.com. If you have questions, please contact us at 800-275-0102 or LeadCareSupport@magellandx.com.
Initiated
2018-08-08
Root cause
Process control
Terminated
2020-05-12
Distribution
Worldwide distribution: US (Nationwide) and countries of: Spain and South Africa.
Recall number
Z-0285-2019
Recall status
Terminated
CDRH record ID
167389
FDA record created
2018-10-26
Recall event number
81074

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗