Open original FDA record ↗Recall record
- FDA identifier
- Z-0286-2011
- Company in FDA record
- GE Healthcare, LLC
- FDA classification date
- Mar 1, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Failure to properly document the CTDI in the Technical Reference or User Manual.
Source facts
- Action
- Planned action(s) to repair defect or to bring product into compliance: GE Healthcare will bring the Technical Reference Manual into compliance by correcting the noncompliance in forward production and the installed base units. This will be accomplished by delivering an "Important Electronic Product Radiation Warning" letter and a TRM addendum which contains the corrected information to customers. This will be carried out via a Field Modification Instruction IFMI 229251 planned to be released in October, 2010 with an estimated completion date of April, 2011. The letter identifies the affected product and describes the defect and related hazards. It also asks customers to refer to the TRM addendum for the scan time accuracy, the helical scan exposure time, and the CTDI adjustment factors for 4 and 8 slices. Contact information is also provided.
- Initiated
- 2010-05-01
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2013-12-05
- Distribution
- Nationwide Distribution
- Recall number
- Z-0286-2011
- Recall status
- Terminated
- CDRH record ID
- 94928
- FDA classified
- 2011-03-01
- Recall event number
- 56922