Open original FDA record ↗Recall record
- FDA identifier
- Z-0286-2021
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Oct 23, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
RX tunnel component (black sheath) may detach from the catheter shaft due to lower bond strength.
Source facts
- Action
- All consignees were sent a copy of the customer notification letter via Federal Express Priority Overnight mail on 09/16/2020. The notification instructs the customer to check their inventory, remove any affected product from their inventory, and return it to BSC. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory. Follow-up communications will be made to the accounts that do not respond to this inquiry. The returned products will be quarantined until an appropriate course of action has been determined. BSC will notify FDA of this decision prior to its implementation.
- Initiated
- 2020-09-16
- Root cause
- Vendor change control
- Distribution
- Worldwide distribution - US Nationwide and International distribution.
- Recall number
- Z-0286-2021
- Recall status
- Open, Classified
- CDRH record ID
- 183804
- FDA record created
- 2020-10-23
- Recall event number
- 86476