Skip to content

Public FDA record

Z-0286-2021: Hurricane RX Biliary Balloon Dilatation Catheter Catalog REF No. M00545890; M00545900; M00545910; M005459

Recall record Z-0286-2021 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0286-2021
Company in FDA record
Boston Scientific Corporation
FDA record creation date
Oct 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

RX tunnel component (black sheath) may detach from the catheter shaft due to lower bond strength.

Source facts

Action
All consignees were sent a copy of the customer notification letter via Federal Express Priority Overnight mail on 09/16/2020. The notification instructs the customer to check their inventory, remove any affected product from their inventory, and return it to BSC. The notification also requests the completion of a Reply Verification Tracking Form as a method of documenting the presence or absence of affected product from their inventory. Follow-up communications will be made to the accounts that do not respond to this inquiry. The returned products will be quarantined until an appropriate course of action has been determined. BSC will notify FDA of this decision prior to its implementation.
Initiated
2020-09-16
Root cause
Vendor change control
Distribution
Worldwide distribution - US Nationwide and International distribution.
Recall number
Z-0286-2021
Recall status
Open, Classified
CDRH record ID
183804
FDA record created
2020-10-23
Recall event number
86476

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.