Skip to content

Public FDA record

Z-0286-2025: epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515

Recall record Z-0286-2025 for Siemens Healthcare Diagnostics Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0286-2025
Company in FDA record
Siemens Healthcare Diagnostics Inc
FDA classification date
Nov 1, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Siemens Healthcare Diagnostics Inc. has confirmed that a negative sodium bias may exist in one lot of epoc BGEM BUN Test Cards approximately 2 months prior to expiration. The observed maximum negative bias from internal testing is -10 mmol/L. No positive bias has been observed. Quality Control (QC) fluid may or may not detect the issue depending on the bias magnitude and the frequency of QC testing. One customer complaint has been associated with this issue. No adverse events have been reported. All affected lots have recently expired before the issuance of the recall.

Source facts

Action
An Urgent Medical Device Correction letter dated September 2024 was sent to customers. Customer Actions " Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " If you are a distributor, please ensure your customers receive this UMDC letter. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " Adverse reactions or quality problems experienced with the use of the affected products listed in the above table may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax. " Please retain this letter with your laboratory records and forward this letter to those who may have received this product. " For any questions, you may call Siemens Healthineers Technical Solutions Center at 877-229-3711.
Initiated
2024-09-24
Root cause
Under Investigation by firm
Distribution
US Nationwide. Canada.
Recall number
Z-0286-2025
Recall status
Open, Classified
CDRH record ID
210439
FDA classified
2024-11-01
Recall event number
95504

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗