Z-0286-2027: Orthoscan VERSA Accessory Cart, REF: 1000-0021 and Versa Mini C-Arm System, REF: 1000-0020

Recall record Z-0286-2027 for ZIEHM-ORTHOSCAN INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0286-2027
Company in FDA record
ZIEHM-ORTHOSCAN INC
FDA classification date
Oct 23, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The accessory cart weldment does not conform to weld dimensional specifications which may result in separation of the monitor arm mounting structure from the cart.

Source facts

Action
On September 11, 2026, Ziehm-Orthoscan issued a Urgent: Medical Device Recall Notification via Email. Ziehm-Orthoscan ask consignees to take the following actions: 1. The device may continue to be used until the corrective action has been completed, provided the user follows the precautions described in this notification and no signs of instability or unintended movement are observed. 2. Users should exercise caution when positioning, moving, or operating the system and monitor assembly, avoid fully extending the display arm until the corrective action has been performed. 3. If unintended movement, instability, loosening, or any other condition related to the issue described in this notification is observed, discontinue use of the affected system and contact Ziehm- Orthoscan immediately for further instructions. 4. Please ensure that a copy of this notification is made available to all users and operators of the affected device. It is recommended that this notification be posted in a visible location near the device until the corrective action has been completed. 5. If the affected device has been transferred, sold, or relocated to another facility or department, please ensure that a copy of this notification is provided to the current user or owner of the device. 6. Please maintain awareness of this notification until the corrective action has been completed and verified by Ziehm-Orthoscan or its authorized representative.
Initiated
2026-09-03
Root cause
Process control
Distribution
Worldwide - US Nationwide distribution in the states of AZ , CA , CO , CT , FL , GA , IL , IN , KS, MA , MD , MI , MN , NC , NE , NV , NY , OH , OR , SC, TX , UT , WA and the countries of Germany , Australia , Spain , Switzerland, Japan.
Recall number
Z-0286-2027
Recall status
Open, Classified
CDRH record ID
223114
FDA classified
2026-10-23
Recall event number
99979

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗