Skip to content

Public FDA record

Z-0287-2011: enpath medical Myopore Sutureless Myocardial Pacing Lead, Bipolar (labeled as Unipolar) Connector, , REF

Recall record Z-0287-2011 for Greatbatch Medical. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0287-2011
Company in FDA record
Greatbatch Medical
FDA classification date
Nov 2, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Myopore Bipolar Epicardial Leads packaging is labeled as unipolar leads. The lead body itself is correctly labeled as bipolar. This could result in a bipolar lead being implanted with a pulse generator operating in unipolar mode.

Source facts

Action
Greatbatch Medical issued an Urgent Medical Device Recall letter dated May 13, 2010 to their distributor. The letter described the problem and product involved, and gave the action required which included: 1) immediately discontinue distribution of the units and return any unused units to Greatbatch Medical. 2) Reconcile medical recordsbetween BSC and Greatbatch Medical to confirm implant data 3) Confirm pulse generator type (bipoloar, unipolar, or unknown) used with each serial number listed Greatbatch will work with customers to contact physicians as necessary regarding patient monitoring. Greatbatch Medical can be contacted at 763-951-8312.
Initiated
2010-05-13
Root cause
Employee error
Terminated
2012-05-10
Distribution
Worldwide Distribution: to USA and Europe, through distributor in MN.
Recall number
Z-0287-2011
Recall status
Terminated
CDRH record ID
93235
FDA classified
2010-11-02
Recall event number
56324

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗