Reason stated in the source
The Myopore Bipolar Epicardial Leads packaging is labeled as unipolar leads. The lead body itself is correctly labeled as bipolar. This could result in a bipolar lead being implanted with a pulse generator operating in unipolar mode.
Source facts
- Action
- Greatbatch Medical issued an Urgent Medical Device Recall letter dated May 13, 2010 to their distributor. The letter described the problem and product involved, and gave the action required which included: 1) immediately discontinue distribution of the units and return any unused units to Greatbatch Medical. 2) Reconcile medical recordsbetween BSC and Greatbatch Medical to confirm implant data 3) Confirm pulse generator type (bipoloar, unipolar, or unknown) used with each serial number listed Greatbatch will work with customers to contact physicians as necessary regarding patient monitoring. Greatbatch Medical can be contacted at 763-951-8312.
- Initiated
- 2010-05-13
- Root cause
- Employee error
- Terminated
- 2012-05-10
- Distribution
- Worldwide Distribution: to USA and Europe, through distributor in MN.