Open original FDA record ↗Recall record
- FDA identifier
- Z-0287-2013
- Company in FDA record
- Dako North America Inc.
- FDA classification date
- Nov 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Dako is recalling the PT Link Waterbath because of a possible electrical hazard. Specifically, if the heater pad under the reagent tank becomes wet, it may present an electrical shock hazard when the lid is open.
Source facts
- Action
- The firm, Dako, sent a "DAKO PT LINK ADVISORY NOTICE" dated May 1, 2008 to customers who purchased the PT Waterbath along with a customer acknowledgement document informing the users of the potential electrical hazard. The notice described the product, problem and actions to be taken. The customers were instructed to follow the attached Safety Instructions; place the enclosed warning labels directly on the instrument as shown in the attached instructions and adhere to the warning labels as described. Customers with questions were instructed to call Dako Products' Technical Support at (800) 424-0021.
- Initiated
- 2008-05-01
- Root cause
- Device Design
- Terminated
- 2012-11-13
- Distribution
- Nationwide distribution: USA including states of: AK, CA, CO, CT, DC, FL, GA, HI, IA, IL IN, KS, KY, MA, MD, MI, MN, MO, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, and WA.
- Recall number
- Z-0287-2013
- Recall status
- Terminated
- CDRH record ID
- 85291
- FDA classified
- 2012-11-13
- Recall event number
- 53336