Heater vent door on neonatal incubator may malfunction.
Source facts
Action
The recalling firm notified domestic consignees by certified letter and international distributors by e-mail dated 10/17/2003. The recall notice was incorporated into a firm''s follow-up letter concerning an on-going product correction of the elevating base on the same neonatal beds. The letter advises of the malfunctioning heater door vent and states that the firm will replace all existing vent doors with doors made of higher temperature rated material. Firm will provide retrofit kits for consignee or service representative installation.
Initiated
2003-10-17
Root cause
Other
Terminated
2005-10-11
Distribution
Products were sold to 365 direct accounts around the US and 58 international distributors.
Recall number
Z-0288-04
Recall status
Terminated
CDRH record ID
29961
FDA classified
2004-07-20
Recall event number
27490